Nano-Cancer® therapy is carried out in a 100 kHz magnetic field applicator (MFH®-300F) which was developed specifically for this form of therapy. The unit has a treatment gap with an adjustable treatment width between 21 and 33 centimeters. The spool current of the 100 kHz oscillating current is continuously adjustable from 100 to 500 amperes. The magnetic alternating field can be adjusted depending on the gap distance, the spool current and the vertical distance from the gap center, from approximately 2 to 18 kA/m. Within the treatment gap, the desired overheating area can be positioned in a circular area with a diameter of nearly 20 cm.

Neither immobilization nor anesthesia of the patient is required to perform thermotherapy. Due to its construction, the alternating magnetic field applicator is universally usable for tumors in all body regions. Temperature measurements during therapy are made in a minimally invasive manner with a fluoroptical measuring system and corresponding display and documentation software. The entire treatment takes approximately two hours and, in addition to the 60-minute therapy period, includes the time for warming up the target area as well as periods for preparing and positioning of the patient.